Configuration and Use Manual
109
Troubleshooting
Op
eratio
n
App
e
ndi
ces
Mai
n
tenan
c
e and
 T
roub
le
sh
oo
tin
g
Com
m
is
si
on
in
g
A021
Unrecognized/ 
Unentered Sensor 
Type
The sensor is recognized 
as a straight tube but the 
K1 value indicates a 
curved tube, or vice versa.
• Check the characterization. Specifically, verify the FCF 
and K1 values. See Section 4.2.
• Check the sensor RTD circuitry. See Section 10.24.
• If the problem persists, contact Micro Motion.
Incorrect Sensor 
Type (K1)
A029
Internal 
Communication 
Failure
Transmitter electronics 
failure
• Cycle power to the flowmeter.
• Contact Micro Motion. 
PIC/Daughterboard 
Communication 
Failure
A030
Hardware/Software 
Incompatible
The loaded software is not 
compatible with the 
programmed board type.
• Contact Micro Motion. 
Incorrect Board 
Type
A031
Undefined
The transmitter is not 
receiving enough power.
• Check power supply to transmitter. See Section 10.14.1.
• If the alarm is appearing and disappearing frequently, 
check the mA Fault Action setting. If it is set to Upscale: 
consider a lower setting for fault level, consider changing 
the fault action to another setting (e.g., Internal Zero), or 
consider increasing the power supply to the transmitter. 
See Section 10.7.
Low Power
A033
Tube Not Full
No signal from LPO or 
RPO, suggesting that 
sensor tubes are not 
vibrating.
• Verify process. Check for air in the flow tubes, tubes not 
filled, foreign material in tubes, or coating in tubes.
Tube Not Full
A100
Primary mA Output 
Saturated
The calculated amount of 
current output is outside of 
the linear range.
Primary mA Output 
Saturated
A101
Primary mA Output 
Fixed
Non-zero HART address 
configured, or user has 
fixed the mA output.
• Check the HART address. If non-zero, enable Loop 
Current Mode parameter. See Section 10.19.
• Exit mA output trim. See Section 7.4.
• Exit mA output loop test. See Section 7.3.
• Check to see if the output has been fixed via digital 
communication. 
Primary mA Output 
Fed
A102
Drive Over-Range
The drive power 
(current/voltage) is at its 
maximum
• Excessive drive gain. See Section 10.23.2.
• Check the sensor circuitry. See Section 10.24.
• If this is the only active alarm, it can be ignored. If desired, 
reconfigure the alarm severity to Ignore (see 
Section 5.3.1).
Drive Overrange
A104
Calibration-In- 
Progress
A calibration procedure is 
in progress.
• Allow the flowmeter to complete calibration.
• For zero calibration procedures, you may abort the 
calibration, set the zero time parameter to a lower value, 
and restart the calibration.
Calibration in 
Progress
A105
Slug Flow
The density has exceeded 
the user-defined slug 
(density) limits.
Slug Flow
Table 10-3
Status alarms and remedies continued
Alarm 
code
Communicator 
Cause
Suggested remedy
ProLink II