SECTION 7—TROUBLESHOOTING GUIDE
26
SECTION 7—TROUBLESHOOTING 
GUIDE
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
Alarm:
Short beeps, long 
pause
Concentrator not 
operating,
power switch On.
Beep......Beep......
Main Power Loss:
1. Power cord not 
plugged in.
1. Insert plug into outlet.
2. No power at outlet.
2. Inspect house circuit 
breakers or fuses. If problem 
recurs, use a different outlet.
3. Tripped circuit 
breaker.
3. Call Service Provider.
Alarm:
Continuous
Concentrator not 
operating, Power 
Switch On. 
Beeeeeeeep....
System Failure:
1. Unit overheating due 
to blocked air intake.
1a. Remove and clean cabinet 
filters.
1b. Move oxygen 
concentrator at least three 
inches away from walls, 
draperies or furniture.
2. Insufficient power at 
outlet.
2. DO NOT use extension 
cords. Move to another 
electrical outlet or circuit.
3. Internal repairs 
required.
3. Call Service Provider.
Alarm:
Rapid
Beep....Beep...
Beep...Beep
Low Flow Alarm:
1. Kinked or blocked 
tubing, cannula or 
humidifier.
1. Inspect for kinks or 
blockages. Correct, clean or 
replace item. Once 
corrected, turn power Off for 
60 seconds and then turn 
power back ON.
2. Flowmeter set at 0.5 
L/min or less.
2. Reset flowmeter to 
prescribed flowrate. If 
prescribed flowrate is below 
0.5 L/min, use the Pediatric 
Flowmeter IRCPF16 (Only 
with Perfecto
2
 5).