SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
18
 WARNING
The fillport cover on the connector fillport and the post 
valve fillport MUST be replaced after filling and whenever 
not in use. If either of the fillport covers are missing 
contact your healthcare provider before using.
17. Place the fillport covers onto the connector fillport (located 
on the compressor) and post valve fillport (located on the 
cylinder) (FIGURE 4.7).
18. Install a CGA870 post valve compatible conserver or 
regulator according to the manufacturer’s instructions.
19. Open the toggle valve by turning counterclockwise and 
examine the conserver or regulator gauge to ensure 
cylinder is full (needle pointing into GREEN area). If the 
cylinder is not full, perform the following:
• If cylinder is not full after second attempt to fill, contact 
the provider.
FIGURE 4.7   Connector Fillport
Connector 
Fillport Cover
Fillport Cover
Toggle 
Valve