SECTION 5—OPERATING INSTRUCTIONS
21
Connect Humidifier (If So Prescribed)
NOTE: For this procedure, refer to FIGURE 5.1,FIGURE 5.2,
FIGURE 5.3 on page 22 and FIGURE 5.4 on page 23.
FIGURE 5.3 on page 22 and FIGURE 5.4 on page 23.
1. Fill humidifier with distilled water to the level indicated
by the manufacturer. (Please contact Invacare for
recommendations).
recommendations).
WARNING
DO NOT overfill humidifier.
Replace the humidifier cap and securely tighten.
DO NOT reverse the oxygen input and output
Replace the humidifier cap and securely tighten.
DO NOT reverse the oxygen input and output
connections. Water from the humidifier bottle will travel
through the cannula back to the patient.
FIGURE 5.1 Filling the Humidifier
2. Place the humidifier bottle in the humidifier
compartment.
FIGURE 5.2 Humidifier Compartment
Humidifier Bottle
Humidifier Bottle