Mintek LV-T Manual Do Utilizador

Página de 172
Stratos LV/LV-T Technical Manual  143 
 
1. Insert the enclosed torque wrench 
through the perforation in the self-
sealing plug at an angle perpendicular 
to the lead connector until it is firmly 
placed in the setscrew. 
2.  Insert the lead connector pin into the 
connector receptacle of the CRT-P 
without bending the lead until the 
connector pin becomes visible behind 
the setscrew.  Hold the connector in 
this position. 
3.  Securely tighten the setscrew of the 
connector clockwise with the torque 
wrench until torque transmission is 
limited by the wrench. 
4.  After retracting the torque wrench, the 
perforation will self-seal.  The proximal 
electrode of bipolar leads is 
automatically connected.  Connect the 
second lead as described above. 
5.  Attach the other ventricular lead to the 
right ventricular port (RV) and the atrial 
lead to the right atrial port (RA) in the 
same manner.