SECTION 1—GENERAL GUIDELINES
9
Perfecto
2
™Series
Radio Frequency Interference
This equipment has been tested and found to comply with 
EMC limits specified by IEC/EN 60601‐1‐2. These limits are 
designed to provide a reasonable protection against 
electromagnetic interference in a typical medical installation.
Other devices may experience interference from even the low 
levels of electromagnetic emissions permitted by the above 
standards. To determine if the emissions from the Perfecto
2
 is 
causing the interference, turn the Perfecto
2
 Off. If the 
interference with the other device(s) stops, then the Perfecto
2
 
is causing the interference. In such rare cases, interference 
may be reduced or corrected by one of the following 
measures:
• Reposition, relocate, or increase the separation 
between the equipment.
• Connect the equipment into an outlet on a circuit 
different from that to which the other device(s) is 
connected.
Avoid creation of any spark near medical oxygen equipment. 
This includes sparks from static electricity created by any type 
of friction.