Invacare 636959 Manual Do Utilizador

Página de 32
4
Part No 636959
FIGURE 1   Medical Suction Unit
Using the Suction Unit
CAUTION
Should fluid be aspirated back into the unit, send your suction unit to an 
authorized service center for a detailed inspection as the vacuum pump may 
have been damaged.
NOTE: For this procedure, refer to FIGURE 1 and FIGURE 2 on page 5.
NOTE: If the unit sustains a severe drop, validity of the gauge must be checked. Inspect if 
the float shut‐off is still functioning properly when the liquid in the collection bottle reaches 
the safe full level.
NOTE: Always transport unit with vacuum adjust knob rotated fully clockwise in case 
unit is dropped.
1. Ensure the suction tubing is connected to the labeled patient connector(outlet) 
on the collection bottle.
2. Ensure that all connections are secure and without leaks.
3. Connect the unit to power.
4. Turn the power switch On (I).
5. Adjust the vacuum level from 0 to 560 mmHg by turning the vacuum adjust 
knob: 
• Clockwise to increase vacuum. 
• Counter‐clockwise to decrease vacuum.
NOTE: Refer to the vacuum gauge while setting the desired level of vacuum. To accurately 
read the gauge, block the patient end of the hose or cap off the collection bottle and allow the 
gauge to reach a stable vacuum reading.
Vacuum 
Adjustment 
Knob
Vacuum Gauge
Collection 
Bottle
Holder
Hydrophobic Filter
Connection 
Tubing
Power Cord
Fuse
Elbow Connector
Lid
On (I) 
Position
Off (O) 
Position
Power 
DETAIL “A”
Connection 
Tubing