SECTION 6—OPERATING INSTRUCTIONS
Invacare XPO
2
18
Location
 WARNING
NEVER block the air openings of the product or place it on a 
soft surface, such as a bed or couch, where the air opening 
may be blocked. Keep the openings free from lint, hair and the 
like.
Keep unit at least three inches away from walls, draperies, 
furniture, and the like. 
Locate and position the XPO
2
 in a well ventilated space so that 
the air intake and the air exhausts are not obstructed. 
Recommendations for Optimal 
Performance
Operating Temperature:
41°F to 95°F (5°C to 35°C)
Relative Humidity:
15% to 60%
Transport/Storage 
Temperature:
-2°F to 140°F (-20°C to 60°C)
Allow unit to warm, or cool, to operating 
temperature range before using.
Electrical:
No extension cords. 
Altitude:
Up to 10,000 ft (3046 m) above sea level.
Tubing and Cannula:
4 ft cannula with a maximum 25 ft of crush 
resistant tubing (DO NOT pinch).
Environment:
Smoke, pollutant and fume free. No 
confined spaces (example: no closets).
Time of Operation:
Up to 24 hours per day when connected to 
AC or DC power adapter.