SECTION 1—GENERAL GUIDELINES
9
Invacare XPO
2
For optimum performance, Invacare recommends that each 
concentrator be on and running for a minimum of 5 minutes. 
Shorter periods of operation may reduce maximum product 
life.
The XPO
2
 is not designed to be used with a humidifier. Use of 
this device with a humidifier may impair performance and/or 
damage the equipment.
When your automobile is turned Off, disconnect the car 
accessory power supply and remove the XPO
2
 from the 
automobile. NEVER allow the XPO
2
 to be stored in a very hot 
or cold automobile or in other similar, high or low, 
temperature environments. Refer to Typical Product 
Parameters on page 14.
DO NOT operate in temperatures below 41° F (5°C) or above 
104° F (40° C) for extended periods of time.
The XPO
2
 MUST be used in an upright position.
Invacare recommends the XPO
2
 not be used in the rain.
Maintenance
The XPO
2
 was specifically designed to minimize routine 
preventive maintenance. Only professionals of the healthcare 
field or persons fully conversant with this process such as 
factory trained personnel should perform preventive 
maintenance or performance adjustments on the oxygen 
concentrator. 
Radio Frequency Interference
This equipment has been tested and found to comply with 
EMC limits specified by IEC/EN 60601‐1‐2. These limits are 
designed to provide a reasonable protection against 
electromagnetic interference in a typical medical installation.