Philips M3500B Manual Do Utilizador

Página de 180
Cleaning Instructions
11-12
Maintaining the HeartStart XLT
Cleaning Instructions
Following are recommendations for cleaning the HeartStart XLT and its asso-
ciated case, cables, etc.
Cleaning the HeartStart XLT
The following cleaning products may be used to clean the exterior surfaces of 
the HeartStart XLT, as well as the battery and data card:
z
Isopropyl alcohol (70% in water)
z
Mild soap and water
z
Chlorine bleach (3% in water)
z
Quaternary ammonium compounds, such as Lysol (10% in water)
When cleaning, be sure to avoid pouring fluids on the device and do not allow 
fluids to penetrate the exterior surfaces of the device. Use of a soft cloth is 
recommended for cleaning the display, to prevent scratching.
CAUTION
The HeartStart XLT may not be autoclaved, ultrasonically cleaned, or 
immersed. Do not use abrasive cleaners or strong solvents such as acetone or 
acetone-based cleaners.
Cleaning the Carrying Case
Clean the carrying case by hand, using a mild soap and water. A brush may be 
used on stubborn spots and a fabric cleaner may be used to remove grease 
spots. Air dry the carrying case when cleaning is complete.
If it is necessary to remove the carrying case for cleaning, refer to directions 
on page 11-15.