Philips M3500B Manual Do Utilizador

Página de 180
Specifications
13-2
Specifications & Safety
Specifications
Defibrillator
Waveform: Biphasic Truncated Exponential. Waveform parameters adjusted 
as a function of patient impedance.
Shock Delivery: Via multifunction defib electrode pads.
Charge Time: Less than 3 seconds to 200 Joules with a new, fully charged 
M3516A SLA battery pack at 25
o
C. Less than 15 seconds to 200 Joules when 
powered by a M3517A or M3518A power module with no battery installed. 
Patient Impedance Range: 25 to 180 Ohm.
Manual Mode
Manual Output Energy (Delivered): 5, 10, 25, 50, 70, 100, 150, 200 
Joules.
Controls: On/Energy Select, Off, Charge/Disarm, Shock, ECG Lead 
Select, SpO
2
, HR Alarms, Manual Mode, Sync, Pacer, Pacer Start/Stop, 
Pacer Rate, Pacer Current, Pacer Mode, Display Contrast, ECG Gain, Vol-
ume, Print Strip, Print Summary, Annotated Mark Event.
Indicators: LCD display for ECG waveform and text prompts, Audio 
alerts, QRS Beeper, Charging Tone, Charge Done Tone, Manual Mode 
LED, Sync LED, Pacer LED, Printer.
Armed Indicators: Charge done tone and available energy indicated on 
display.
Energy Selection: Front Panel "1" key.
Charge Control: Front Panel "2" key.
Shock Control: Front Panel "3" key.
Synchronizer: SYNC message appears on the monitor and is annotated on 
the printer (if printing while in Sync Mode). An audible beep sounds with 
each detected R-wave, while a tick mark on the monitor and printed strip 
indicate the discharge points. Synchronizer delay is less than 60 msec from 
peak R-wave to peak current of the defibrillation discharge.