Safety Considerations
13-19
13
Safety Considerations
The following general warnings and cautions apply to use of the HeartStart 
XLT. Additional warning and cautions specific to a particular feature are pro-
vided in the appropriate section. 
WARNING
The HeartStart XLT is not intended to be deployed in settings or situations that pro-
mote use by untrained personnel. Operation by untrained personnel can result in 
injury or death.
WARNING
Remain attentive to the patient during the delivery of therapy. Delay in delivering a 
shock may result in a rhythm that was analyzed as shockable converting spontane-
ously to non-shockable and could result in inappropriate delivery of a shock. 
WARNING
Use only the multifunction defib electrode pads, battery, and accessories listed in 
“Supplies & Accessories” o
n page 11-18. Substitutions may cause the HeartStart 
XLT to function improperly.
WARNING
Use multifunction defib electrode pads prior to their expiration date. Discard pads 
after use. Do not reuse pads.
WARNING
In AED Mode, the multifunction defib electrode pads must be in the anterior-anterior 
position as shown on the packaging. The HeartStart XLT was not designed to assess 
data acquired from pads in an anterior-posterior position.