Philips M3500B Manual Do Utilizador

Página de 180
Disabling Sync Mode
7-4
Performing Synchronized Cardioversion
Delivering Additional Synchronized Shocks
If additional synchronized shocks are indicated, make sure Sync Mode is still 
enabled and repeat steps 2-5. In its default configuration, the HeartStart XLT 
remains in Sync Mode after a shock is delivered, as indicated by the message 
6\QF on the display and the lighted green LED next to 
The HeartStart XLT can be configured to exit Sync Mode after each shock is 
delivered.
Disabling Sync Mode
To disable Sync Mode, press 
.  The green LED next to 
 goes 
out and the message 
6\QF is no longer displayed. Sync Mode is also disabled 
when you exit Manual Mode.
Sync
Sync
Sync