Reminders
Remove any medicine patches and residual adhesive from the 
patient’s chest before applying the pads.
Do not place the pads directly over an implanted pacemaker or 
defibrillator. A noticeable lump with a surgical scar should 
indicate the position of an implanted device.
Do not allow the pads to contact other electrodes or metal 
parts that are in contact with the patient.
If the pads do not stick well, check that the pads adhesive has not 
dried out. Each pad has a layer of adhesive gel. If the gel is not 
sticky to the touch, replace the pads with a new set. (For ease of 
handling, the pad is designed with a non-gel area around the 
connector cable.)
Keep the patient still and keep any movement around the patient 
to a minimum during rhythm analysis. Do not touch the patient 
or the pads while the Caution light is on solid or flashing. If the 
FRx is unable to analyze due to electrical “noise” (artifact), it will 
tell you to stop all movement and remind you not to touch the 
patient. If the artifact continues for more than 30 seconds, the 
FRx will pause briefly to allow you to deal with the source of the 
noise, then resume analysis.
The FRx will only deliver a shock if the flashing orange Shock 
button is pressed when the instruction is given. If the Shock 
button is not pressed within 30 seconds after the instruction, the 
FRx will disarm itself, and (for the first CPR interval) give a 
reminder to make sure emergency medical services have been 
called, then begin a CPR interval. This is designed to minimize 
interruption of CPR and help ensure ongoing patient support.
While waiting for you to press the Shock button, the FRx will 
continue to analyze the heart rhythm. If the patient’s rhythm 
changes before you press the Shock button, and a shock is no 
longer needed, the defibrillator will disarm and tell you a shock is 
not advised.
If for any reason you want to turn off the defibrillator during a 
use, you can press the On/Off button 
– holding it down for at 
least one second 
– to return the device to standby mode.