SECTION 3—CLEANING AND DISINFECTING
24
Cleaning the Compressor
CAUTION
Unplug unit before cleaning. DO NOT submerge in water to 
clean.
DO NOT clean using abrasive cleaners.
1. Clean the casing of the main unit daily by using a soft cloth 
moistened with water or a mild detergent.
2. Wipe the casing and immediately dry using a soft clean 
cloth.
Cleaning the Mouthpiece
1. Boil the mouthpiece in water for a maximum of five 
minutes or autoclave at a maximum of 121° C for ten 
minutes.
2. Allow the mouthpiece to air dry.
3. Once dry, the mouthpiece may be stored inside the 
nebulizer top.