SECTION 6—TROUBLESHOOTING
21
SECTION 6—TROUBLESHOOTING
NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare.
SYMPTOMS
FAULTS
SOLUTION
Noisy or dry sound from pivots.
Needs lubrication.
Electric actuator fails to lift when button is
pressed.
pressed.
Hand-control or actuator
connector loose.
connector loose.
Battery low.
RED emergency stop button
pressed IN.
pressed IN.
Battery not connected properly
to control box.
to control box.
The connecting terminals are
damaged.
damaged.
Electric actuator in need of ser-
vice or load is too high.
vice or load is too high.
Rotate RED emergency stop button
CLOCKWISE until it pops out.
CLOCKWISE until it pops out.
Reconnect the battery to the control box.
Refer to Primary Emergency Release on
page 13.
Refer to Primary Emergency Release on
page 13.
Replace the battery pack. Refer to Primary
Emergency Release on page 13.
Emergency Release on page 13.
Unusual noise from lift arm actuator.
Actuator is worn or damaged
or spindle is bent.
or spindle is bent.
Lift arms will not lower in uppermost position.
Lift arms require a minimum
weight load to lower from the
uppermost position.
weight load to lower from the
uppermost position.
Pull down slightly on the lift arms.
Legs don’t open and close properly.
Leg actuators may be worn or
damaged or spindle is bent.
damaged or spindle is bent.