OPTIONS Ramp/Taper 
Mode
 
Vista
 
basic 
Instructions 
for 
Use 
     13/25 
950787 Rev H 02/03
 
 
 
To Clear Ramp/Taper Mode 
1.  Press OPTS until display shows RAMP TAPER MODE. 
2. Press 
CLR. 
3.  Press YES to clear all the data in Ramp/Taper Mode. 
4.  Press ON and continue to program a new Ramp/Taper program. 
 
To Exit Ramp/Taper Mode 
1.  When infusion is complete, press   
to stop the KVO. 
2. Press 
EXIT. 
3.  Press END to return to Continuous Mode. 
 
Fast Taper  
The infusion can be interrupted to begin an early taper down process during the 
Ramp/Taper Mode Infusion. 
 
1.  Stop the infusion. 
2. Press 
FAST. 
3.  Enter the FAST Taper Time in hours and minutes.  
4. Press 
OK. 
5.  Press    