IDEAL INDUSTRIES Meter Manual Do Utilizador

Página de 23
Page 8 
 
Required Equipment 
          Required equipment is listed in Table B. If the recommended  
          models are not available, equipment with equivalent specifications may be used. 
          Repairs or servicing should be performed only by qualified personnel. 
 
Table B. Required Equipment 
Equipment 
Required Characteristics 
Recommended 
Model
 
Calibrator 
AC Voltage Range: 0-750V ac 
Accuracy: ±0.07% (Basic) 
Frequency Range: 40 ~ 100KHz 
Accuracy: ± 2% 
 
DC Voltage Range: 0-1000V dc 
Accuracy: ±0.006% (Basic) 
 
Current Range: 0 ~ 10A 
Accuracy:  
AC (40Hz to 1KHz): ±0.08% (Basic) 
DC: ± 0.02% (Basic) 
 
Frequency Source: 
 5.00Hz ~ 1.0000MHz 
Accuracy: ±0.001%  
 
Amplitude:  
0.5V p-p ~ 1.0V p-p (square wave) 
Accuracy: ± 5% 
 
Ω range: 
1Ω ~ 2G 
Accuracy: ±0.03% (Basic) 
 
Capacitance Range: 
1PF ~ 40mF 
Accuracy: ±0.10% (Basic) 
 
Temperature Range: 
-200 ~ 1200 
Accuracy: ±0.3 (Basic) 
Fluke 5500 or 
Wavetek 9100 
Calibrator or  
equivalent 
 
 
Form Number TM61633-5 
Rev 5 Dec 2006