Testing the Capacitance Function (cont’d) 
      d)  For each input, compare the readings on the Meter display to display readings in  
            Table 7. The meter selects the proper range automatically. Each measurement takes  
            about one second per range. 5mF will taken about 15 seconds. 
3.   If the display reading falls outside of the range shown in Table 7, the Meter does not 
      meet specification. 
 
Table 7. Capacitance Test 
Input Readings 
Step Capacitance 
61-633 
61-635 
3.600nF 
3.512 to 3.688 
3.548 to 3.652 
36.00nF 
35.12 to 36.88 
35.48 to 36.52 
360.0nF 
351.2 to 368.8 
354.8 to 365.2 
4 3600μF 
3.512 to 3.688 
3.548 to 3.652 
5 36.00μF 
34.76 to 37.24 
35.12 to 36.88 
6 360.0μF 
347.6 to 372.4 
351.2 to 368.8 
1.000mF 
.9410 to 1.059 
.9510 to 1.049 
 
 
Checking the Diode Test Function 
To check the diode test function, do the following: 
1.   Connect the Calibrator to the VΩ and COM inputs on the Meter. 
2.   Turn the rotary switch to 
3.   Apply 3.000V. 
      The meter display should read approx. 3.000V dc. 
4.   Press “Blue” key to test 
 function. 
5.   Apply a 50Ω resistor to meter, the built-in beeper alarms. 
 
 
Testing the Milliamp (mA) Function 
To verify the accuracy of DC and AC current measurement functions, do  
The following: 
1.    Connect the Calibrator to the mA and COM inputs on the Meter. 
2.    Turn the rotary switch to mA. 
3.    Apply the inputs for steps 1-2 in Table 8. 
4.    For each input, compare the readings on the Meter display to the display readings in  
       Table 8. 
 
Table 8. DC mA Test 
Input Readings 
Step Current 
61-633 
61-635 
36.000rnA 
35.780 to 36.220 
35.888 to 36.112 
360.00mA 
357.80 to 362.20 
358.88 to 361.12 
 
5.    Press the blue button on the Meter to toggle to AC measurement function. 
6.    Apply the inputs for steps 1-4 in Table 9. 
7.    For each input, compare the readings on the Meter display to the display readings in 
       Table 9. 
8.    If the display reading falls outside of the range shown in the Tables, the meter does not 
       meet specification. 
 