•  Click on the Sound, video and game controllers listing. 
•  If the LBA-PC frame grabber was detected, and the device driver was not loaded you will see a 
category called Unknown in the edit box listing of the Device Manager.  If this occurs, double 
click on the Unknown icon.  You should see an entry called PCI Card.  This indicates that the 
device driver was not properly installed.  To correct this problem, re-install the LBA-PC software. 
•  If you did not see the Unknown icon, then Windows did not detect the LBA-PC frame grabber 
card.  Check to see that your frame grabber card is properly installed.  Sometimes PCI card slots 
are defective, so you might also try another slot if all else seems to be in order. 
If this device driver is not found, the LBA-PC program will complete loading, but the software will be 
forced to the Off-Line mode.  This means you can only view files and perform Post Processing of data 
from and/or to data files. 
 
Device Drive vX.XX detected. 
LBA-PC requires vY.YY. 
LBA-PC set to Off-Line mode. 
 
If your current device driver version does not match the current LBA-PC version you will get this 
message.  This may occur if you have upgraded your LBA-PC application, but continue to use an old 
device driver.  Normally this should not happen, as the installation will install a new device driver 
version with each upgrade.  The LBA-PC application program will complete loading, but the software will 
be forced to the Off-Line mode.  This means you can only view files and perform Post Processing of 
data from and/or to data files. 
 
Unable to load LCA program file. 
File Not Found. 
 
This message will precede the following error message if the lcaXXX.exo file is not found or is not in the 
same path as the LBA-PC application. 
 
LBA-PC frame grabber 
detected, but cannot initialized. 
LBA-PC set to Off-Line mode. 
 
The LBA-PC frame grabber contains a programmable LCA device.  If this device fails to program during 
application initialization, the above error message will occur.  Contact the Spiricon Service Department 
for assistance.  The LBA-PC application program will complete loading, but the software will be forced 
to the Off-Line mode.  This means you can only view files and perform Post Processing of data from 
and/or to data files. 
 
Not Enough Memory for Frame Capture. 
LBA-PC set to Off-Line mode. 