10.4 Command Formats and Responses 
Commands to the LBA-PC will not normally generate a response back to the host controller, unless the 
command: 
•  Changes remote/local mode 
•  Is a query 
•  Initiates an operation that produces results 
•  Generates an error 
 
 
10.4.1     IEEE 488.1 Command Support 
The LBA-PC in combination with an appropriate National Instruments GPIB board conforms to the 
IEEE 488.1 standard according to the following Subset Codes: 
 
 
SH1, AH1, T6, L4, SR1, RL2, PP0, DC1, DT0, E2, C0 
 
These Subset Codes are defined in the IEEE 488.1 standard and describe interface capabilities of the 
LBA-PC. 
The following multiple line interface message commands are supported by the LBA-PC: 
 
GTL 
Go To Local (0x01).  When the LBA-PC detects the GTL message it returns 
to local command mode. 
 
LLO    - 
Local Lockout (0x11).  When the LBA-PC detects the LLO message it enters 
what is called the lockout state.  This is equivalent to entering the operator 
password in the Options | Password menu item.  While in the lockout state 
access is denied to many of the LBA-PC menu items.  When the LBA-PC 
exits the lockout state access to all menu items is restored. 
 
MLA 
My Listen Address (0x20 to 0x3e). 
 
MSA 
My Secondary Address (0x60 to 0x7e). 
 
MTA 
My Talk Address (0x40 to 0x5e). 
 
PPC 
Parallel Poll Configure (0x05).  Handled by GPIB board and NI-488.2M. 
 
PPD 
Parallel Poll Disable (0x70 to 0x7e).  Handled by GPIB board and NI-
488.2M. 
 
PPE 
Parallel Poll Disable (0x60 to 0x6e).  Handled by GPIB board and NI-
488.2M. 
 
PPU 
Parallel Poll Unconfigure (0x15).  Handled by GPIB board and NI-488.2M. 
 
SPD 
Serial Poll Disable (0x19). 
 
SPE 
Serial Poll Enable (0x18). 
 UN 
Unlisten 
(0x3f). 
 UNT 
Untalk 
(0x5f). 
 
Operator’s Manual 
 
LBA-PC 
 
Doc. No. 10654-001, Rev 4.10 
179
The following IEEE 488.1 commands are not supported: