TOOL
.
   
VI Basic Operations 
 
3    Measuring Two-point Distance or Entering Information—TOOL Menu 
- 72 - 
3) 
Draw a measurement line. 
Click the mouse button at the start point of the object to be measured and hold down the 
mouse button while you drag and release the mouse button at the end point, to draw a 
line. The length of the measurement line will be displayed in the vicinity of it.   
Each measurement line is assigned numbers 1 to 9. The number of measurement lines 
drawn in the same color is limited to nine. To change colors, click another color icon, so 
that measurement lines drawn in that color will be assigned numbers separately from 
those in other colors. 
To redraw a line, use    Undo    or    Redo    to cancel or redo your most recent action. Up to 
10 recent actions can be retrieved. 
Use    Clear    to erase all the content you’ve now entered. All lines you’ve drawn will be 
erased unless you have pasted. 
NOTE:  If you draw a measurement line in the middle of    Undo    or    Redo , recent 
measurement lines drawn from that point in time cannot be canceled. 
4) 
To save a measurement line to the overlay of the image, press    Paste    to fix after 
drawing it. 
5) Press    to return to the menu in which you previously were. Any measurement lines 
you’ve created will be erased unless you pasted. Once the content is fixed as valid, you 
cannot erase it unless you erase the entire overlay. 
NOTE: 
The display information is interlocked with the menu size. Small characters 
are displayed when you are showing a small menu, and large characters are 
displayed when you are showing a large menu. 
Use this function when the units "nm," "µm," "mm," and "cm" are displayed. 
If other units are displayed, you need to make the calibration setting once 
again. 
If the microscope connected to your system is the Nikon 80i/90i microscope 
with motorized or intelligent nosepiece, note that each calibration value call 
number is linked to the hole number of the nosepiece. This means you 
cannot change calibration values by selecting    s1    through    s7 . 
Calibration values are automatically changed when you change the 
objective.