VI Basic Operations 
 
TOOL 
 
 
3    Measuring Two-point Distance or Entering Information—TOOL Menu 
- 75 - 
 
3.5.2 
Using a Scale to Measure the Subject Size 
 
 
 
 
 
 
 
 
You can display a scale on the screen and use it to measure an approximate size of the subject. 
 
1) 
Select the calibration value you want to use from    s1    through    s7 . 
2) 
Press    Scale , and a scale matched to the selected calibration value will be displayed.     
3) 
Press    Scale    again to erase the displayed scale. 
 
A scale normally is displayed on only the screen, and is not drawn in the saved or printed image. 
If you select the Scale check box on the REC menu, two scales, one vertical and one horizontal, 
will be drawn in the lower right part of the saved or printed image. 
 
 
 
NOTE: 
Depending on the calibration value (such as when the screen-full length is 
calibrated as being equal to 1 mm), the system may not be able to display 
values. In such a case, the value is displayed as “######” and you must 
change the unit of the calibration value from “1 mm” to “1000 µm,” etc. 
Photographed images whose image size is 2,560 x 1,920 cannot be saved or 
printed with scales written over the image. If 2,560 x 1,920 is selected for 
the image size, it will be automatically reduced to 1,280 x 960. 
Use this function when the units "nm," "µm," "mm," and "cm" are displayed. 
If other units are displayed, you need to make the calibration setting once 
again. 
If you perform an electronic zoom while using this function, the screen 
display will reflect the zoom magnification. 
If you change the optical zoom of the digital imaging head while using this 
function with the Nikon 80i/90i microscope connected to your system, the 
screen display will reflect the zoom magnification. 
If you change the objective while using this function, and if the microscope 
connected to your system is the Nikon 80i/90i microscope using a motorized 
nosepiece or intelligent nosepiece, the screen display will change 
accordingly. 
If the microscope connected to your system is the Nikon 80i/90i microscope 
using a motorized nosepiece or intelligent nosepiece, note that each 
calibration value call number is linked to the hole number of the nosepiece. 
This means you cannot change calibration values by selecting  s1  through  
  s7 . Calibration values are automatically changed when you change the 
objective.