Supplementary instruction manual V 1.2 – 07/2002
5.2 Adjusting the pre-millers in the
calibrating device
Preparatory steps
• Connect mains power supply (20) to miller cali-
bration device (18).
• Connect mains power supply (20) to mains: the
light bulb is switched on.
• Place magnifier (16) onto calibration device.
• Loosen regulating screws (6), (7) and (11) in
miller holder (2) until they are flush with the
inner surfaces of the miller holder.
If the screws protrude from the inner surfaces
of the miller holder (2),  spring (1) and miller (3)
cannot be inserted into the correct position.
Twisted  or  deformed  springs  for  the
miller holders must be exchanged!
Inserting the miller into the miller holder
• Insert spring (1) straight from above into miller
holder (2) - the two upper edges of the spring
must be flush with the upper surface of the
miller holder.
• Carefully insert miller (3) from the front - the
letter  ‘V’ must be visible.
When  the  miller  is  inserted  into  the
holder  the  letter  ‘V’  must  be  visible.  -
That way, the milling edge is pointing
into the right direction.
• Insert the miller (3) completely - it must fit
against the rear and lower inner surface of the
miller holder (2) and be flush with spring (1).
• Insert screw (4) and washer (5) and tighten
slightly to fasten miller (3) .
Fig. 5.1
20
18
16
Fig. 5.2
6
6
11
7
1
3
2
5
4
5. Operation