Operating The T703 over the Serial I/O Ports 
Teledyne API T703 Calibrator Operation Manual 
118  
 
7.3. MULTIDROP RS-232 SET UP 
When the RS-232 Multidrop option is installed, the instrument designated as last in the chain must be 
terminated. This requires installing a shunt between two pins on the multidrop printed circuit assembly (PCA) 
inside the instrument. Step-by-step instructions follow. 
Note that because the RS-232 Multidrop option uses both the RS232 and COM2 DB9 connectors on the 
analyzer’s rear panel to connect the chain of instruments, COM2 port is no longer available for separate RS-232 
or RS-485 operation. 
 
 
 
 
 
 
 
 
 
To install shunt in the last analyzer: 
1. With NO power to the instrument, remove its top cover and lay the rear panel open for access to the 
multidrop PCA, which is seated on the CPU. 
2.  On the multidrop PCAs JP2 connector, use the shunt provided to jumper Pins 21 
 22 as indicated in 
 
  
Figure 7-3: 
Multidrop/LVDS PCA Seated on CPU 
 
CAUTION – Risk of Instrument Damage and Warranty Invalidation 
Printed circuit assemblies (PCAs) are sensitive to electro-static discharges too 
small to be felt by the human nervous system.  Damage resulting from failure to use 
ESD protection when working with electronic assemblies will void the instrument 
warranty.   
See A Primer on Electro-Static Discharge section in this manual for more information on 
preventing ESD damage. 