T703 Calibration and Verification 
Teledyne API T703 Calibrator Operation Manual 
130  
 
 
Figure 8-2: 
Set up for Verifying Optional O
3
 Photometer Using an External O
3
 Generator 
 
 
NOTE 
The manifolds as shown in the above drawing are oriented to simplify the drawing.   
All unused ports should be capped. 
A Minimum of 1.1 LPM is required for the external zero air source 
 
8.1.1.1. Calibration Manifold Exhaust/Vent Line 
The manifold’s excess gas should be vented to a suitable vent outside of the room.  This vent should be  of large 
enough internal diameter to avoid any appreciable pressure drop, and it must be located sufficiently downstream 
of the output ports to assure that no ambient air enters the manifold due to eddy currents or back diffusion.  
NOTE 
It is recommended that the calibration manifold’s exhaust vent have a minimum internal diameter of 3/8 
inch and a maximum length of 3 meters (or 10 feet) 