T703 Calibration and Verification 
Teledyne API T703 Calibrator Operation Manual 
142  
 
8.3. T703 GAS PRESSURE SENSOR CALIBRATION 
The T703 Calibrator has two sensors that monitor the pressure of the gases flowing through the instrument.  The 
data collected by these sensors is used to compensate the final concentration calculations for changes in 
atmospheric pressure and is stored in the CPU’s memory as test functions: 
Table 8-1:  T703 Pressure Sensors 
SENSOR 
ASSOCIATED 
TEST FUNCTION 
UNITS 
PRESSURE MONITOR  
MEASUREMENT POINT 
Regulator Pressure Sensor 
 
REG PRESSURE 
PSIG 
Capped fitting on backside of regulator 
assembly.  See Figure 8-4 
Photometer Sample Gas 
Pressure Sensor 
PHOTO SPRESS 
IN-HG-A 
Use monitor to measure ambient 
atmospheric pressure at the calibrator’s 
location.   
 
8.3.1.1. Gas Pressure Sensor Calibration Set Up 
The procedures described in this section require an independent, calibrated pressure meter/monitor be attached 
at the following location.   
PHOTOMETER
 PRESSURE SENSOR
O
3
GEN / PHOTOMETER 
PRESSURE / FLOW SENSOR PCA
O
3
GAS INPUT 
PRESSURE SENSOR
O
3
FLO
SENS
O
R
O
3
Generato
r A
ssem
b
ly
O
3
GEN
E
RA
TO
R
Flo
w
 C
ont
ro
(10
0 cm
3
)
REF/MEAS
Valve
On Back Panel
Chassis
GAS OUTPUT MANIFOLD
PHOTOMETER 
OUTLET
TO ANALYZER
VENT
TO ANALYZER
DRY AIR 
IN
ZERO AIR 
IN
PHOTOMETER 
INLET
EXHAUST
PHOTOMETER 
ZERO OUT
PHOTOMETER 
ZERO IN
PHOTOMETER BENCH
PUMP
Flow Control 
(800 cm
3
)
INTERNAL
VENT
Flo
w
 Co
nt
rol 
(5.0 lp
m)
Pressure 
Regulator
CHA
R
COA
L
 
SC
RUB
B
ER
Filter
PUMP
Flo
w
 Co
nt
rol 
(1.0
 LP
M)
Pressure 
Monitor
 
Figure 8-3: 
Pressure Calibration Monitor Points 
 
 
 
07223B DCN6378