Teledyne API T703 Calibrator Operation Manual
T703 Calibration and Verification
147
8.4.2. CALIBRATING THE OUTPUT GAS FLOW
8.4.2.1. Output Gas Flow Set Up
The procedure described in this section requires an independent, calibrated flow meter/monitor and the following
set up:
set up:
PHOTOMETER
PRESSURE SENSOR
O
3
GEN / PHOTOMETER
PRESSURE / FLOW SENSOR PCA
O
3
GAS INPUT
PRESSURE SENSOR
O
3
FL
O
O
W
SE
NSOR
NSOR
O
3
G
e
ne
ra
to
r Asse
mb
ly
ly
O
3
GENE
RAT
OR
REF/MEAS
Valve
On Back Panel
Chassis
GAS OUTPUT MANIFOLD
PHOTOMETER
OUTLET
TO ANALYZER
VENT
TO ANALYZER
DRY AIR
IN
ZERO AIR
IN
PHOTOMETER
INLET
EXHAUST
PHOTOMETER
ZERO OUT
PHOTOMETER
ZERO IN
PHOTOMETER BENCH
OFF
INTERNAL
VENT
Flow
Co
ntr
ol
ntr
ol
(5
.0
lp
m)
m)
Pressure
Regulator
CH
A
A
R
CO
A
A
L
SC
R
R
U
B
B
E
R
Filter
PUMP
Flow
Co
ntr
ol
ntr
ol
(1
.0
LPM)
FLOW
METER
CAP
CA
P
P
CA
P
P
Fl
ow C
o
n
trol
(1
0
0 c
m
3
/m
in)
in)
Flow Control
(800 cm
(800 cm
3
/min)
Figure 8-5:
Output Flow Calibration Monitor Point
07223B DCN6378