Theory of Operation 
Teledyne API T703 Calibrator Operation Manual 
156  
 
 
Power 
Connections 
for DC 
Heaters 
Status LED’s  
(D2 through D16) 
DC Power Supply 
Test Points 
Watchdog  
Status LED (D1) 
I
2
C
 Connector 
DC  
Valve Control 
Drivers 
AC Power  
IN 
DC Power 
Distribution 
Connectors 
DC Valve & 
Photometer 
Pump Control 
Connector 
AC Power 
OUT to Dry 
Air Pump 
AC Pump 
Configuration 
Plug 
Dry Air AC Pump 
Control Relay 
DC Valve & 
Photometer  
Pump  
Control 
 
Figure 9-4: 
Relay Board PCA with AC Relay Retainer Removed 
This version of the Relay PCA includes one AC relay that controls the AC-powered Dry Air (zero air) pump and 
A plastic insulating safety shield covers the remaining empty AC Relay sockets. 
 
CAUTION 
NEVER REMOVE THIS SAFETY SHIELD WHILE THE INSTRUMENT IS PLUGGED IN AND 
TURNED ON.  THE CONTACTS OF THE AC RELAY SOCKETS BENEATH THE SHIELD 
CARRY HIGH AC VOLTAGES EVEN WHEN NO RELAYS ARE PRESENT 
 
9.2.3.1. Valve Control 
The relay board also hosts two valve driver IC's, each of which can drive up four valves.  In the T703, the relay 
PCA controls only those valves associated with the O
3
 generator and photometer options.   
07223B DCN6378