Teledyne MODEL T703 Manual Do Utilizador

Página de 287
General Troubleshooting & Repair 
Teledyne API T703 Calibrator Operation Manual 
206  
 
11.4.9.4. Control Inputs 
 
Table 11-11: 
T703 Control Input Pin Assignments and Corresponding Signal I/O Functions 
CONNECTOR 
INPUT 
CORRESPONDING I/O SIGNAL 
Top A 
CONTROL_IN_1 
Top B 
CONTROL_IN_2 
Top C 
CONTROL_IN_3 
Top D 
CONTROL_IN_4 
Top E 
CONTROL_IN_5 
Top F 
CONTROL_IN_6 
Bottom G 
CONTROL_IN_7 
Bottom H 
CONTROL_IN_8 
Bottom I 
CONTROL_IN_9 
Bottom J 
CONTROL_IN_10 
Bottom K 
CONTROL_IN_11 
Bottom L 
CONTROL_IN_12 
 
 
The control input bits can be tested by applying a trigger voltage to an input and watching changes in the status 
of the associated function under the SIGNAL I/O submenu: 
EXAMPLE: to test the “A” control input: 
1. Under 
the 
DIAG SIGNAL I/O menu (See Section11.1.3), scroll through the inputs and outputs until 
you get to the output named 0) CONTROL_IN_1.  
2.  Connect a jumper from the
 
+
” pin on the appropriate connector to the “U” on the same connector.   
3.  Connect a second jumper from the “” pin on the connector to the “A” pin.   
4.  The status of 0) CONTROL_IN_1 should change to read “ON”.   
 
 
 
 
07223B DCN6378