General Troubleshooting & Repair 
  
215 
11.8. REPAIR PROCEDURES 
11.8.1. REPAIRING SAMPLE FLOW CONTROL ASSEMBLY 
The critical flow orifice is housed in the flow control assembly (Teledyne API part number: 001760400) located 
on the top of the optical bench.  A sintered filter protects the jewel orifice so it is unusual for the orifice to need 
replacing, but if it does, or the filter needs replacement please use the following procedure (see the Spare Parts 
list in Appendix B for part numbers and kits): 
1.  Turn off power to the calibrator. 
2.  Locate the assembly to be repaired, see Figure, 3–3. 
3.  Disconnect the pneumatic connection from the flow assembly. 
4.  Remove the fitting and the components as shown in the exploded view in Figure 11.6.   
5.  Replace the o-rings (p/n OR000001) and the sintered filter (p/n FL000001). 
6.  If replacing the critical flow orifice itself (P/N 000941000), make sure that the side with the colored 
window (usually red) is facing downstream to the gas flow. 
7.  Apply new Teflon
®
 tape to the male connector threads 
8.  Re-assemble in reverse order.   
 
Pneumatic Connector, Male 1/8”
(P/N FT_70
Spring
(P/N HW_20)
Sintered Filter
(P/N FL_01)
Critical Flow Orifice
(P/N 000941000) 
  Make sure it is placed with the  
  jewel down) 
O-Ring
(P/N OR_01)
Purge Housing
(P/N 000850000)
 
Figure 11-5: 
Critical Flow Restrictor Assembly Disassembly 
 