Teledyne API T703 Calibrator Operation Manual 
Getting Started 
 
43 
3.4.5. OPERATING MODES FOR THE O
3
 GENERATOR 
The O
3
 generator can be set to operate in three different modes: 
3.4.5.1. CNST (CONSTANT) 
In this mode, the O
3
 output of the generator is based on a single, constant, drive voltage.  There is no Feedback 
loop control by the T703’s CPU in this mode. 
3.4.5.2. REF (REFERENCE) 
The O
3
 control loop will use the reference detector’s measurement of the O
3
 generator’s UV lamp as input.  This 
mode does not use the photometer to control the ozone generator. 
3.4.5.3. BNCH (BENCH) 
The O
3
 concentration control loop will use the photometer’s O
3
 measurement as input.   
THIS IS THE DEFAULT AND MOST COMMON MODE OF OPERATION. 
This setting will be the default mode of the T703 calibrator and will be mused whenever the calibrator is using 
the GENERATE  AUTO command or the GENERATE sequence step to create a calibration mixture.  When 
either the GENERATE  MAN command or the MANUAL sequence step is active, the local O
3
 generator mode 
(chosen during when the command/step is programmed) will take precedence. 
3.4.6. SETTING THE O
3
 GENERATOR MODE 
To select a default O
3
 generator mode, press:   
 