Operating the T703 Calibrator 
Teledyne API T703 Calibrator Operation Manual 
54    
6.1. TEST FUNCTIONS 
A variety of TEST FUNCTIONS are available for viewing at the front panel whenever the calibrator is at the 
MAIN MENU.  These measurements provide information about the present operating status of the calibrator and 
are useful during troubleshooting (see Section 11).  Table 6-1 lists the available TEST functions. 
To view these TEST functions, press <TST or TST> from the main menu and the Test Functions will scroll at 
the top center of the display. 
 
Table 6-1:  Test Functions Defined 
TEST MEASUREMENT 
DESCRIPTION 
ACT=GENERATE XX PPB O3 
Actual concentration being generated, computed from 
real-time inputs. 
TARG=GENERATE XX PPB O3 
Target concentration to generate. 
OUT FLW=X.XXX LPM 
Output flow rate (computed from regulator pressure). 
REG PRESS=XX.X PSIG 
Regulator pressure. 
BOX TMP=XX.X ºC 
Internal chassis temperature. 
OGEN REF=XXXX.X MV 
O
3
 generator reference detector reading. 
O3GENDRV=XXXX.X MV 
O
3
 generator lamp drive output. 
O3LAMPTMP=XX.X ºC 
O
3
 generator lamp temperature. 
PH MEAS=XXXX.X MV 
Photometer detector measure reading. 
PH REF=XXXX.X MV 
Photometer detector reference reading. 
PH FLW=X.XXX LPM 
Photometer sample flow rate. 
PH LTEMP=XX.X ºC 
Photometer lamp temperature. 
PH SPRES=XX.X IN-HG-A 
Photometer sample pressure. 
PH STEMP=XX.X ºC 
Photometer sample temperature. 
PH SLOPE=X.XXX 
Photometer slope computed during zero/span bench 
calibration. 
PH OFFST=X.X PPB 
Photometer offset computed during zero/span bench 
calibration. 
PHOTO STABIL=X.X PPB 
Photometer concentration stability (standard deviation of 
25 bench concentration samples taken 10 seconds apart). 
TEST=XXXX.X MV 
Value output to TEST_OUTPUT analog output, selected 
with TEST_CHAN_ID variable. 
TIME=HH:MM:SS 
Current instrument time of day clock.  (24 hour format) 
 