Teledyne MODEL T703 Manual Do Utilizador
Operating the T703 Calibrator
Teledyne API T703 Calibrator Operation Manual
54
6.1. TEST FUNCTIONS
A variety of TEST FUNCTIONS are available for viewing at the front panel whenever the calibrator is at the
MAIN MENU. These measurements provide information about the present operating status of the calibrator and
are useful during troubleshooting (see Section 11). Table 6-1 lists the available TEST functions.
MAIN MENU. These measurements provide information about the present operating status of the calibrator and
are useful during troubleshooting (see Section 11). Table 6-1 lists the available TEST functions.
To view these TEST functions, press <TST or TST> from the main menu and the Test Functions will scroll at
the top center of the display.
the top center of the display.
Table 6-1: Test Functions Defined
TEST MEASUREMENT
DESCRIPTION
ACT=GENERATE XX PPB O3
Actual concentration being generated, computed from
real-time inputs.
real-time inputs.
TARG=GENERATE XX PPB O3
Target concentration to generate.
OUT FLW=X.XXX LPM
Output flow rate (computed from regulator pressure).
REG PRESS=XX.X PSIG
Regulator pressure.
BOX TMP=XX.X ºC
Internal chassis temperature.
OGEN REF=XXXX.X MV
O
3
generator reference detector reading.
O3GENDRV=XXXX.X MV
O
3
generator lamp drive output.
O3LAMPTMP=XX.X ºC
O
3
generator lamp temperature.
PH MEAS=XXXX.X MV
Photometer detector measure reading.
PH REF=XXXX.X MV
Photometer detector reference reading.
PH FLW=X.XXX LPM
Photometer sample flow rate.
PH LTEMP=XX.X ºC
Photometer lamp temperature.
PH SPRES=XX.X IN-HG-A
Photometer sample pressure.
PH STEMP=XX.X ºC
Photometer sample temperature.
PH SLOPE=X.XXX
Photometer slope computed during zero/span bench
calibration.
calibration.
PH OFFST=X.X PPB
Photometer offset computed during zero/span bench
calibration.
calibration.
PHOTO STABIL=X.X PPB
Photometer concentration stability (standard deviation of
25 bench concentration samples taken 10 seconds apart).
25 bench concentration samples taken 10 seconds apart).
TEST=XXXX.X MV
Value output to TEST_OUTPUT analog output, selected
with TEST_CHAN_ID variable.
with TEST_CHAN_ID variable.
TIME=HH:MM:SS
Current instrument time of day clock. (24 hour format)
07223B DCN6378