Operating the T703 Calibrator 
Teledyne API T703 Calibrator Operation Manual 
66    
6.5.1.6. Setting Up Control Outputs for a Sequence 
The T703 calibrator’s control outputs allow the entire sequence to be triggered from an external source.  This 
feature allows the calibrator to control devices that accept logic-level digital inputs, such as programmable logic 
controllers (PLCs), dataloggers, or digital relays/valve drivers.   
They can be used as: 
  12 separate ON/OFF switches assigned to separate calibration sequences, or; 
  A 12-bit wide bus allowing the user to define activation codes for up to 4095 separate calibration 
sequences.  
They can be set to: 
  Be active whenever a particular calibration sequence is operating, or; 
  Activate/deactivate as individual steps within a calibration sequence are run See Section 6.5.2.5). 
To assign a CC OUTPUT pattern / code to a particular sequence, press. 
STANDBY            ACT =STANDBY
<TST
TST>
GEN STBY SEQ
SETUP
Make sure that the 
calibrator is in standby 
mode.
SETUP X.X        
PRIMARY SETUP MENU
O3
SEQ
CFG CLK PASS
MORE
EXIT
SETUP X.X        
SEQUENCE CONFIGURATION
EDIT
PRINT
EXIT
SETUP X.X               
END OF SEQUENCES
INS
PRNT
EXIT
SETUP X.X               
1) SEQ [NAME][X] STEPS
PREV NEXT
INS
DEL EDIT
PRNT
EXIT
SETUP X.X     NAME:0
SET>
EDIT
EXIT
Deletes the sequence shown
in the message field
Edits the sequence shown 
in the message field
Scrolls back and forth between 
existing sequences
SETUP X.X    CC OUTPUT:DISABLED
<SET SET>
EDIT
EXIT
Continue pressing SET> until ...
This display only appears if there are no sequences currently 
programmed into the 
calibrator
.
OTHERWISE ...
SETUP X.X     CC OUTPUT:[0]00000000000
<CH CH>
  [0]
ENTER EXIT
EXIT discards the 
new setting
ENTR accepts the 
new setting
SETUP X.X     CC OUTPUT ENABLE:OFF
OFF
ENTER EXIT
 Toggle to turn 
the CC output 
ON/OFF
EXIT discards the 
new setting
ENTR accepts the 
new setting
 Toggle to turn the selected bit ON/OFF (0 or 1).  
Each bit shown on the display represents one of the control 
output pins located on the calibrator rear panel (see Figure 3-2),
The left most bit is Bit 1, the next bit to the right, bit 2, 
progressing rightward to bit 12 (see Figure 3-10 for connector pin 
assignments)
Moves the 
cursor one 
character left or 
right.
 