Teledyne MODEL T703 Manual Do Utilizador

Página de 287
Teledyne API T703 Calibrator Operation Manual 
Operating the T703 Calibrator 
 
69 
6.5.2.1. The GENERATE Step 
 This step operates and is programmed similarly to the GENERATE  AUTO.  
At the end of the programming sequence, the T703 firmware will automatically insert a DURATION step that 
needs to be defined. 
To insert a GENERATE step into a sequence, press: 
INSERT STEP Submenu
SETUP X.X          INSERT STEP: GENERATE
PREV NEXT
ENTR
EXIT
Starting at the INSERT STEPS
Submenu
EXIT discards the new 
target concentration
ENTR accepts the new 
target concentration
SETUP X.X         GENERATE:0.0 PPB  O3
0
.0
0
0
PCT
O3
ENTR EXIT
SETUP X.X 
GENERATE:ZERO
 ZERO
ENTR
EXIT
SETUP X.X          3) DURATION : 10.0 MIN
PREV NEXT
INS
DEL EDIT
EXIT
Toggle to switch 
between ZERO 
AIR and O
3
modes.
SETUP X.X         GENERATE:0.0 PPB  O3
0
0
0
.0
PPB
O3
ENTR EXIT
Toggle these 
buttons to set the 
target 
concentration.
Toggle to scroll 
through the 
available units of 
measure
SETUP X.X 
DURATION: 1.0 MIN
0
0
1
.0
ENTR
EXIT
Toggle these 
buttons to set 
DURATION of this 
step
 
07223B DCN6378