6.8. SETUP  PASS 
The T703 provides password protection of the calibration and setup functions to prevent unauthorized 
adjustments.  When the passwords have been enabled in the PASS menu item, the system will prompt the user 
for a password anytime a password-protected function is requested. 
There are three levels of password protection, which correspond to operator, maintenance and configuration 
functions.  Each level allows access to all of the functions in the previous level. 
Table 6-6:  Password Levels 
PASSWORD 
LEVEL 
MENU ACCESS ALLOWED 
No password 
Operator 
All functions of the MAIN menu: TEST, GEN, initiate SEQ , MSG, CLR  
101 Maintenance 
Access to Primary and Secondary Setup Menus except for VARS & DIAG 
818 Configuration 
Secondary SETUP Submenus VARS and DIAG 
 
To enable or disable passwords, press: 
 
07223B DCN6378