Operating the T703 Calibrator
91
6.9.2.3. Manual Calibration of the TEST Channel Configured for Voltage Ranges
For highest accuracy, the voltages of the analog outputs can be manually calibrated.
NOTE:
The menu for manually adjusting the analog output signal level will only appear if the AUTO-CAL feature
is turned off for the channel being adjusted (see Section6.9.2.1)
Calibration is performed with a voltmeter connected across the output terminals (See Figure 6-2) and by
changing the actual output signal level using the front panel keys in 100, 10 or 1 count increments.
changing the actual output signal level using the front panel keys in 100, 10 or 1 count increments.
V
+DC Gnd
Figure 6-3:
Setup for Calibrating the TEST CHANNEL
Table 6-10: Voltage Tolerances for the TEST CHANNEL Calibration
FULL
SCALE
ZERO
TOLERANCE
SPAN VOLTAGE
SPAN
TOLERANCE
MINIMUM
ADJUSTMENT
(1 count)
0.1 VDC
±0.0005V
90 mV
±0.001V
0.02 mV
1 VDC
±0.001V
900 mV
±0.001V
0.24 mV
5 VDC
±0.002V
4500 mV
±0.003V
1.22 mV
10 VDC
±0.004V
4500 mV
±0.006V
2.44 mV