Teledyne T100 Manual Do Utilizador

Página de 422
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Data Acquisition System (DAS) and APICOM 
 
 
161 
 
When editing the data channels, the top line of the display indicates some of the 
configuration parameters. For example, the display line: 
0) CONC: ATIMER, 4, 800
 
represents to the following configuration: 
CHANNEL NUMBER.: 0 
NAME: CONC 
TRIGGER EVENT: ATIMER 
PARAMETERS: Four parameters are included in this channel 
EVENT: This channel is set up to record 800 data points. 
To edit the name of a data channel, refer to Figure 7-3, then press: 
 
 
 
 
 
 
 
 
SETUP X.X      
NAME:CONC
 
 
   
  C        O        N        C        1        - 
ENTR 
EXIT
 
ENTR
 accepts the new string 
and returns to the previous 
menu. 
EXIT
 ignores the new string 
and returns to the previous 
menu. 
Press each as many times as needed to cycle through the 
character set until desired character appears: 
0-9, A-Z, space ’  ~ ! 
 # $ % ^ & * ( ) - _ = +[ ] { } < >\ | ; : , . / ?  
 
 
 
SETUP X.X      
NAME:CONC1
 
 
    <SET    SET>
  EDIT 
   PRINT 
 
EXIT
 
 
Figure 7-4: DAS – Editing Data Channel Name 
 
06807C DCN6650