Teledyne T100 Manual Do Utilizador

Página de 422
Data Acquisition System (DAS) and APICOM 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
164 
To configure the parameters for a specific data parameter, follow the instructions as 
shown in Figure 7-6, then press:  
 
SETUP X.X     
0) PARAM=CONC1,   MODE=AVG
 
 
PREV  NEXT
    
        INS    DEL   EDIT                   EXIT 
 
SETUP X.X     
PARAMETERS:CONC1
 
 
            SET>  EDIT
 
    
EXIT 
 
SETUP X.X    
SAMPLE MODE:AVG
 
 
<SET   SET>
    
EDIT
    
EXIT 
 
SETUP X.X     
PARAMETERS: CONC1
 
 
PREV  NEXT 
 
 
ENTR  
EXIT
 
 
Cycle through list of available 
Parameters. 
ENTR
 accepts the new 
setting and returns to the 
previous menu. 
EXIT
 ignores the new setting 
and returns to the previous 
SETUP X.X    
SAMPLE MODE: AVG
 
 
INST  AVG  SDEV  MIN   MAX
 
    ENTR    EXIT 
 
Press any button for the desired mode 
SETUP X.X   
PRECISION: 1
 
 
<SET   SET>
    
EDIT
      
EXIT 
 
SETUP X.X   
PRECISION: 1
 
    1 
 
 
 
     ENTR    EXIT 
 
Set for 0-4 
SETUP X.X   
STORE NUM. SAMPLES: OFF
 
 
<SET    
        
EDIT
    
EXIT 
 
SETUP X.X   
STORE NUM. SAMPLES: OFF
 
 
 OFF   
    
 
 
 
    ENTR    EXIT
 
 
Turn ON or OFF 
<SET
 Returns to 
previous 
Functions 
 
Figure 7-7: DAS – Configuring Parameters for a Specific Data Parameter 
06807C DCN6650