Calibrations should be carried out at the field monitoring site. The Analyzer should be in 
operation for at least several hours (preferably overnight) before calibration so that it is 
fully warmed up and its operation has stabilized. During the calibration, the T100 should 
be in the CAL mode, and therefore sample the test atmosphere through all components 
used during normal ambient sampling and through as much of the ambient air inlet 
system as is practicable. If the instrument will be used in more than one range (i.e. 
DUAL OR  AUTO ranges), it should be calibrated separately on each applicable range. 
Calibration documentation should be maintained with each analyzer and also in a central 
backup file. 
Table 10-1: 
Activity Matrix for Calibration Equipment & Supplies 
EQUIPMENT & 
SUPPLIES 
ACCEPTANCE LIMITS 
FREQUENCY AND METHOD
OF MEASUREMENT 
ACTION IF 
REQUIREMENTS ARE NOT
MET 
Recorder 
Compatible with output 
signal of analyzer; min. 
chart width of 150 mm (6 in) 
is recommended 
Check upon receipt 
Return equipment to supplier
Sample line and 
manifold 
Constructed of PTFE or 
glass 
Check upon receipt 
Return equipment to supplier
Calibration equipment  Meets guidelines of 
reference 1 and Section 
2.3.2 (Q.A. Handbook) 
Refer to Section 2.3.9 (Q.A. 
Handbook) 
Return equipment/ supplies 
to supplier or take corrective 
action 
Working standard SO
2
 
cylinder gas or SO
2
 
permeation tube 
Traceable to NIST-SRM 
meets limits in traceability 
protocol for accuracy and 
stability (refer to Section 
2.0.7, Q.A. Handbook) 
Analyzed against NIST-SRM; 
refer to protocol in Section 
2.0.7, Q.A. Handbook 
Obtain new working 
standard and check for 
traceability 
Zero air 
Clean dry ambient air, free 
of contaminants that cause 
detectable response with 
the SO
2
 analyzer. 
Refer to Section 2.9.2 (Q.A. 
Handbook) 
Obtain air from another 
source or regenerate. 
Record form 
Develop standard forms 
N/A 
Revise forms as appropriate 
Audit equipment 
Must not be the same as 
used for calibration 
System must be checked out 
against known standards 
Locate problem and correct 
or return to supplier 
 