Instruction Manual 
760007-A 
July 2003 
Rosemount Analytical Inc.    A Division of Emerson Process Management                                                 Description and Specifications       1-3 
Model NGA2000 PMD 
 
1-5 SPECIFICATIONS
1
 
a. General 
Measurement Species................... Oxygen 
Ranges .......................................... 0 to 100% oxygen; four fullscale selections, including suppressed 
zero ranges in 1% increments 
Repeatability..................................
±1% of fullscale (at constant temperature) 
Minimum Detectable Level ............ 0.01% oxygen 
Noise ............................................. <1% of fullscale, peak-to-peak; <
±1% for suppressed ranges 
Linearity .........................................
±1% of fullscale 
Response Time ............................. 0 to 90% of fullscale in 20 seconds (
±2 seconds) 
Drift (Zero and Span)..................... <
±1% of fullscale/24 hours, <±2% of fullscale/week at constant 
temperature; 
<
±2% of fullscale/24 hours, <±4% of fullscale/week of range for 99 
to 100% (at constant temperature) 
Effect of Temperature.................... <
±1% of fullscale over any 10°C interval for rate of change no 
greater than 10
°C per hour 
Environment .................................. Location - Class B controlled, indoor, non-hazardous 
Ambient Temperature.................... 0 to 45
°C (32 to 113°F) 
Effect of Flow................................. <
±1% of range when sample flow rate is changed by 20 ml/min. 
Power Requirements..................... 24 VDC 
±5%, 50 W max.; ripple and noise: <100 mV peak-to-peak; 
line and load regulations: <
±1% 
 
b. Sample 
Temperature .................................. Non-flammable;: 10 to 66
°C (50 to 150°F) 
Flow Rate ...................................... 800 to 1400 ml/min. 
Exhaust Pressure .......................... -345 to 690 hPa-gauge (-5 to 10 psig) 
Particles......................................... filtered to <2 microns 
Dewpoint........................................ below 43
°C (110°F), no entrained liquid 
Materials in Contact with Sample .. Glass, 316 stainless steel, titanium, Paliney No. 7, epoxy resin, Vi-
ton A, platinum, nickel, rhodium and MgF
2
 
Sample Humidity ........................... non-condensing at ambient temperatures 
 
c. Physical 
Case Classification ........................ General purpose for installation in weather-protected areas 
Dimensions.................................... See Outline and Mounting Dimensions, Figure 2-6 on page 2-7
 
Weight ........................................... 8 kg (17.6 lbs. 
Mounting........................................ Inside a Platform or custom-installed in a panel 
Maximum Length of LON Cable .... 1600 m (1 mile) between Analyzer Module and Platform 
 
 
 
 
                                                      
1
 See the Platform manual for specifications regarding Platform related components.