3-6 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA2000 PMD
pressed range. Table 3-3 below shows rec-
ommended zero and span standard gases for
ommended zero and span standard gases for
suppressed range oxygen indication in an Ar-
gon background.
gon background.
RANGE % OXYGEN
RECOMMENDED ZERO
STANDARD GAS
RECOMMENDED SPAN
STANDARD GAS
0 to 1
Nitrogen
0.9% O
2
, balance N
2
0 to 2.5
Nitrogen
2.3% O
2
, balance N
2
0 to 5
Nitrogen
4.5% O
2
, balance N
2
0 to 10
Nitrogen
9% O
2
, balance N
2
0 to 25
Nitrogen
Air (20.93% O
2
)
0 to 50
Nitrogen
45% O
2
, balance N
2
0 to 100
Nitrogen
100% O
2
Table 3-2. Calibration Range for Various Zero Based Operating Ranges
RANGE % OXYGEN
RECOMMENDED ZERO STANDARD
GAS
RECOMMENDED SPAN
STANDARD GAS
50 to 100
50 - 50.5% O
2
Balance AR
100% O
2
70 – 100%
70 – 70.5% O
2
Balance AR
100% O
2
98 – 100%
98 – 98.5% O
2
Balance AR
100% O
2
99 – 100%
99 – 99.5% O
2
Balance AR
100% O
2
Table 3-3. Calibration Range for Various Suppressed Range Operations
If the user is unable to calibrate the Analyzer
Module (i.e., when ZERO or SPAN is initiated,
nothing happens), a possible solution relates
to the use of an incorrect gas for zeroing or
spanning (e.g., using a high concentration gas
to zero or a zero gas to span the Analyzer
Module). Recalibrating with the appropriate
gas(es) WILL NOT correct the problem be-
cause the ZERO OFFSET or SPAN FACTOR
has been set to an extreme value in the proc-
ess.
Module (i.e., when ZERO or SPAN is initiated,
nothing happens), a possible solution relates
to the use of an incorrect gas for zeroing or
spanning (e.g., using a high concentration gas
to zero or a zero gas to span the Analyzer
Module). Recalibrating with the appropriate
gas(es) WILL NOT correct the problem be-
cause the ZERO OFFSET or SPAN FACTOR
has been set to an extreme value in the proc-
ess.
To remedy the problem, do the following:
1. Select the following from the Main Menu:
Expert Controls and Setup (enter security
code if necessary), Analyzer Module
Setup, and Calibration Parameters.
Setup, and Calibration Parameters.
2. Using the down arrow, select Zero
Ranges, press ENTER and, using the
up/down arrows, toggle to SEPARATE.
Do the same for the Span Ranges selec-
tion. Do not press ESCAPE at any time
unless retention of prior settings is de-
sired.
up/down arrows, toggle to SEPARATE.
Do the same for the Span Ranges selec-
tion. Do not press ESCAPE at any time
unless retention of prior settings is de-
sired.
3. Return to the Main Menu (HOME) and
make the following selections: Expert
Controls and Setup (enter security code if
necessary), Expert Analyzer Controls,
CAL softkey, FACTORS softkey, and
Controls and Setup (enter security code if
necessary), Expert Analyzer Controls,
CAL softkey, FACTORS softkey, and