Rosemount Analytical Inc.    A Division of Emerson Process Management                                                  Maintenance and Service    4-1 
Model NGA2000 PMD 
 
 
PRESSURE 
COMPENSATION 
BOARD 
NETWORK INPUT  
DETECTOR 
HOUSING 
CASE 
TEMPERATURE 
SENSOR 
FLOW SENSOR 
COMPUTER 
ANALYSIS BOARD 
MODULE BOARD 
SECTION 4 
 MAINTENANCE AND SERVICE 
WARNING 
QUALIFIED PERSONNEL 
This equipment should not be adjusted or 
repaired by anyone except properly quali-
fied service personnel. 
4-1 OVERVIEW 
PMD Analyzer components that may require 
replacement include: 
•  All printed circuit board 
•  Thermal fuse inside Detector 
•  Case temperature sensor 
• Flow 
sensor 
• Power 
fuse 
• Detector 
• Module 
fan 
The LED bi-cell assembly source required ad-
justment (rotation) anytime the Detector is 
disassembled.  Refer to Figure 4-3 on page 4-
3 for locations of these components. 
4-2  PRINTED CIRCUIT BOARD REPLACEMENT 
Refer to Figure 4-1 belowfor locations of the 
Signal, Microprocessor, Power Supply and 
(optional) Pressure Compensation Boards. 
All boards are secured to a side of the ana-
lyzer module that folds out while interconnec-
tion wiring is still in place.  Remove the 
securing screws and fold out the entire panel. 
To remove individual boards on the fold-out 
panel, label and unplug all interconnection 
wiring, and remove securing hardware.  Re-
verse this procedure for installation. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 4-1.  PMD Module – Major Components