Sierra Monitor Corporation 201 Folheto

Página de 2
 
2007 by Sierra Monitor Corporation Printed in U.S.A.
Model 203 - Hydrogen Sulfide
The Model 203 can be used in such typical applica-
tions as wastewater treatment plants, laboratories and
chemical plants.
Model 206 - Carbon Monoxide
The Model 206 Carbon Monoxide Gas Sensor Module
can be used to provide an  early warning of Carbon
Monoxide accumulation in tanks, laboratories, chemi-
cal plants, parking garages, vehicle test facilities and
more.
SPECIFICATIONS
SENSOR TYPE:
Solid-State Metal Oxide Semiconductor
RANGE:
(at 50% relative humidity)
  Model 201
300 - 2,000 ppm Hydrogen
500 - 10,000 ppm Methane
  Model 203
10-50 ppm Hydrogen Sulfide (H
2
S)
  Model 206
50 - 500 ppm Carbon Monoxide (CO)
RESPONSE TIME:
  Model 201
Less than 30 seconds
  Model 203
For 50 ppm alarm, if 50 ppm H2S is present:
1-4 minutes, if >250ppm H2S is present:
30-90 seconds
  Model 206
Less than 30 seconds
PERIODIC
 MAINTENANCE:
None, other than routine calibration
OPERATING
 TEMPERATURE
 RANGE:
-4°F TO 158°F
(-20°C TO 70°C)
INPUT:
9 - 24 VDC (+ 1V), 250 mA
OUTPUT:
0.5 amp dry contact, 100 VDC, 130 VAC
Nominal 5 volts DC (TTL), source 25mA.
SAFE and ALARM signals.
INDICATORS:
Internal LED for calibration
ENCLOSURE:
Class I, Division 1, Groups C, D
ENCLOSURE
Cast aluminum for protection against
 MATERIAL:
galvanic corrosion
SIZE:
6.75 x 4.0 x 3.5 in
(17.1 x 10.2 x 6.0 cm)
WEIGHT:
24 oz (678 g)
APPROVAL:
UL 60950
Sierra Monitor Corporation  1991 Tarob Ct. Milpitas, CA  95035  (408) 262-6611  800-727-4377  FAX: (408) 262-9042
Visit our Web site at:  www.sierramonitor.com         E-Mail: sierra@sierramonitor.com
Rev. B 8/23/07
ORDERING INFORMATION
Combustible Gas
201-00
Hydrogen Sulfide
203-00
Carbon Monoxide
206-00