Teledyne API – Model T300/T300M CO Analyzer 
Setup Menu 
107 
5.4.4. RANGE UNITS 
The T300/T300M can display concentrations in parts per million (10
6
 mols per mol, 
PPM) or milligrams per cubic meter (mg/m
3
, MG).  Changing units affects all of the 
display, COMM port and DAS values for all reporting ranges regardless of the 
analyzer’s range mode.  To change the concentration units:  
 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
In order to avoid a reference temperature bias, the analyzer must be 
recalibrated after every change in reporting units. 
 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
Concentrations displayed in mg/m3 and ug/m3 use 0
° C 
and 760 mmHg for 
Standard Temperature and Pressure (STP).  Consult your local regulations 
for the STP used by your agency. 
 (Example: US EPA uses 25 
o
C as the reference temperature). 
Once the Units of Measurement have been changed from volumetric (ppb or 
ppm) to mass units (µg/m3 or mg/m3) the analyzer MUST be recalibrated, as 
the “expected span values” previously in effect will no longer be valid.  
Simply entering new expected span values without running the entire 
calibration routine IS NOT sufficient. 
This will also counteract any discrepancies between STP definitions.
 