Setup Menu 
Teledyne API – Model T300/T300M CO Analyzer 
114 
5.8. SETUP VARS: VARIABLES SETUP AND DEFINITION 
The T300/T300M has several user-adjustable software variables, which define certain 
operational parameters.  Usually, these variables are automatically set by the 
instrument’s firmware, but can be manually redefined using the VARS menu.   
The following table lists all variables that are available within the 818 password 
protected level.  See Appendix A-2 for a detailed listing of all of the T300/T300M 
variables that are accessible through the remote interface. 
Table 5-3:  Variable Names (VARS) 
NO. VARIABLE 
DESCRIPTION 
ALLOWED 
VALUES 
VARS 
DEFAULT 
VALUES 
DAS_HOLD_OFF 
Changes the Internal Data Acquisition System (DAS) 
HOLDOFF timer (Section 7.1.11). 
No data is stored in the DAS channels during situations 
when the software considers the data to be questionable 
such as during warm up of just after the instrument returns 
from one of its calibration mode to SAMPLE Mode.   
May be set for 
intervals 
between  
0.5 – 20 min 
 
15 min. 
CONC_PRECISION 
Allows the user to set the number of significant digits to the 
right of the decimal point display of concentration and stability 
values. 
AUTO, 1, 2,  
3, 4 
AUTO 
DYN_ZERO
 1  
Dynamic zero automatically adjusts offset and slope of the 
CO response when performing a zero point calibration during 
an AutoCal (see Section 9.4). 
ON/OFF 
OFF 
DYN_SPAN 
1  
Dynamic span automatically adjusts the offsets and slopes of 
the CO response when performing a slope calibration during 
an AutoCal  (see Section 9.4)
ON/OFF 
OFF 
CLOCK_ADJ
 
Adjusts the speed of the analyzer’s clock.  Choose the + 
sign if the clock is too slow, choose the - sign if the clock is 
too fast. 
-60 to +60 
s/day 
0 sec 
STABIL_GAS
Selects which gas measurement is displayed when the STABIL
test function is selected. 
CO; CO
2
 & O
2
 
CO 
1
 
Use
 of the DYN_ZERO and DYN_SPAN features are not allowed for applications requiring EPA equivalency. 
2
 
 This VARS only appears if either the optional O
2
 or CO
2
 sensors are installed. 
 
 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
There are more VARS available when using the password, 929, for 
configuration. Use caution when pressing any buttons while in this setup. 
Any changes made may alter the performance of the instrument or cause 
the instrument to not function properly. Note that if there is an accidental 
change to a setup parameter, press EXIT to discard the changes. 
06864B DCN6314