Setup Menu 
Teledyne API – Model T300/T300M CO Analyzer 
122 
5.9.3.1. ANALOG OUTPUT VOLTAGE / CURRENT RANGE SELECTION 
In its standard configuration, each of the analog outputs is set to output a 0–5 VDC 
signals.  Several other output ranges are available.  Each range has is usable from -5% to 
+ 5% of the rated span.   
Table 5-6:  Analog Output Voltage Ranges 
RANGE NAME 
RANGE SPAN 
MINIMUM OUTPUT 
MAXIMUM OUTPUT 
0.1V 
0-100 mVDC 
-5 mVDC 
105 mVDC 
1V 
0-1 VDC 
-0.05 VDC 
1.05 VDC 
5V 
0-5 VDC 
-0.25 VDC 
5.25 VDC 
10V 
0-10 VDC 
-0.5 VDC 
10.5 VDC 
  The default offset for all VDC ranges is 0-5 VDC. 
CURR 
0-20 mA 
0 mA 
20 mA 
  While these are the physical limits of the current loop modules, typical applications use 2-20 or 4-20 mA for the lower and upper 
limits.  Please specify desired range when ordering this option. 
 
The default offset for all current ranges is 0 mA.
 
 
Current outputs are available only on A1-A3.
 
 
 
06864B DCN6314