Setup Menu 
Teledyne API – Model T300/T300M CO Analyzer 
126 
5.9.3.4. AUTOMATIC CALIBRATION OF THE ANALOG OUTPUTS 
To calibrate the outputs as a group with the AOUTS CALIBRATION command, select 
the ANALOG I/O CONFIGURATION submenu (see Figure 5-3) then press: 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
Before performing this procedure, make sure that the AUTO CAL for each 
analog output is enabled. (See Section 5.9.3.3). 
 
  
 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
Manual calibration should be used for any analog output set for a 0.1V 
output range or in cases where the outputs must be closely matched to 
the characteristics of the recording device. 
 