Teledyne API – Model T300/T300M CO Analyzer 
Communications Setup and Operation 
157 
instruments are regarded as slaves of the host computer.  The remote instruments are 
unaware that they are connected to a multidrop bus and never initiate Hessen protocol 
messages.  They only respond to commands from the host computer and only when they 
receive a command containing their own unique ID number. 
The Hessen protocol is designed to accomplish two things: to obtain the status of remote 
instruments, including the concentrations of all the gases measured; and to place remote 
instruments into zero or span calibration or measure mode.  API’s implementation 
supports both of these principal features.   
The Hessen protocol is not well defined, therefore while API’s application is completely 
compatible with the protocol itself, it may be different from implementations by other 
companies. 
Note 
The following sections describe the basics for setting up your instrument 
to operate over a Hessen Protocol network.   
For more detailed information as well as a list of host computer 
commands and examples of command and response message syntax, 
download the Manual Addendum for Hessen Protocol from the Teledyne 
API web site: http://www.teledyne-api.com/manuals/. 
6.7.2.1. HESSEN COMM PORT CONFIGURATION 
Hessen protocol requires the communication parameters of the T300/T300M Analyzer’s 
COMM ports to be set differently than the standard configuration as shown in Table 6-4. 
Table 6-4: 
RS-232 Communication Parameters for Hessen Protocol 
PARAMETER STANDARD 
HESSEN 
Baud Rate 
300 – 19200 
1200 
Data Bits 
Stop Bits 
Parity 
 
None Even 
Duplex Full 
Half 
To change the baud rate of the T300/T300M’s COMM ports, see Section 6.2.2. 
To change the rest of the COMM port parameters listed in Table 6-4, see Sectio6.2 an
Table 6-1. 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
Ensure that the communication parameters of the host computer are also 
properly set. 
 
Note 
The instrument software has a 200 ms latency before it responds to 
commands issued by the host computer. This latency should present no 
06864B DCN6314