Getting Started 
79 
 
 
3.4.4. INITIAL CALIBRATION 
To perform the following calibration you must have sources for zero air and span gas 
available for input into the sample port on the back of the analyzer.  See Section 3.3.2 
for instructions for connecting these gas sources. 
The initial calibration should be carried out using the same reporting range set up as 
used during the analyzer’s factory calibration.  This will allow you to compare your 
calibration results to the factory calibration as listed on the Final Test and Validation 
Data Sheet
.  
If both available DAS parameters for a specific gas type are being reported via the 
instruments analog outputs e.g.  CONC1 and CONC2 when the DUAL range mode is 
activated, separate calibrations should be carried out for each parameter.   
 Use 
the 
LOW button when calibrating for CONC1 (equivalent to RANGE1). 
 Use 
the 
HIGH button when calibrating for CONC2 (equivalent to RANGE2). 
NOTE 
The following procedure assumes that the instrument does not have any 
of the available Valve Options installed. See Section 9.3 for instructions 
for calibrating instruments that have valve options. 